%A VN Fishman %A B Linclau %A DP Curran %A KV Somayajula %J Journal of the American Society for Mass Spectrometry %T Tris (perfluoroalkylethyl)silyl alkyl amines as calibration standards for electron ionization mass spectrometry in the mass range of 100-3000 Da %X A new fluorinated compound mixture has been developed for the calibration of the double focusing mass spectrometer in the mass range of 100-3000 Da in the positive electron ionization (EI) mode. Current calibration standards for EI have either limited mass range [perfluorotributylamine (PFTBA), perfluorokerosene (PFK), s-triazines (TRIS)] or poor peak intensities with significant chemical background in the instrument for several days (perfluoroalkyl phosphazine). The newly synthesized fluorinated silyl alkyl amines mixture is proposed as a reference/calibration standard for EI-MS. This standard produced abundant parent and fragment ions across the entire mass range without any memory effect. Copyright ? 2001 American Society for Mass Spectrometry. %N 9 %P 1050 - 1054 %V 12 %D 2001 %R 10.1016/S1044-0305(01)00284-7 %L pittir20560